Stranded wire; 0.5 to 1.5 mm – Yokogawa ADMAG AXF User Manual

Page 12

Advertising
background image

IM 01E20C01-01E

4-2

4. WIRING

20 (0.8)

10.5 (0.4)

AXFC

L (Specified Dimensions)

150

Ϯ

5

8 (0.3) max.

150

Ϯ

5

70 (2.76)

60 (2.36)

90 (3.54)

(5.9)

(5.9)

55 (2.17)

90 (3.54)

50 (1.97)

25 (0.98)

SA

SB

A

C

B

A

C

B

8(0.3) max.

White Black Red

White Black

Red

F0402.EPS

Unit : mm

(approx. inch)

On the
converter
side

On the
flowtube
side

Figure 4.2.2 Treatment of Dedicated Signal Cables

CAUTION

• As crimp terminals A, B, SA, SB and C have

their own electrical potentials, securely insulate
them so as not to come in contact with one
another.

• To prevent a shield from coming in contact with

another shield or the case, cover each shield
with a vinyl tube or wrap it in vinyl tape.

NOTE

Conductors A and B carry the signal from the
electrodes, and C is at the potential of the liquid
(signal common). Shields SA and SB are kept
at the same potentials as the individual elec-
trodes (these are actively driven shields.) This is
done to reduce the effect of the distributed
capacitance of the cable at long cable length.
Note that, since the signals from the individual
electrodes are impedance converted inside the
converter, errors will result if they come in
contact with any other component. Great care
must be taken in the cable end treatment.

(2) Excitation Cable/Power Cable/Output

Cable

JIS C3401 control cable equivalent
JIS C3312 power cable equivalent
14 AWG Belden 8720 equivalent

Outer Diameter:

With no gland option;

6.5 to 12 mm (0.26 to 0.47 in.)

With gland options EG, EU and EW;

• Excitation cable;

10.5 or 11.5 mm (0.41 or 0.45 in.)

• Power and output cable;

7.5 to 12 mm (0.30 to 0.47 in.)

With gland options EP;

6 to 12 mm (0.24 to 0.47 in.)

Nominal Cross Section:

Single wire; 0.5 to 2.5 mm

2

Stranded wire; 0.5 to 1.5 mm

2

In case of power cable, Green/Yellow covered conduc-
tor shall be used only for connection to PROTECTIVE
CONDUCTOR TERMINALS. Conform to IEC227,
IEC245 or equivalent national authorization.

On the converter side

85 (3.35)

EX1

EX2

EX1

EX2

85 (3.35)

On the flowtube side

Unit : mm

(approx. inch)

F0403.EPS

Crimp Terminal

Figure 4.2.3 End Treatment of Excitation Cable

NOTE

• For excitation and power cables, always use a

crimp terminal with an insulation cover.

• Use crimp tools from the manufacturer of the

crimp terminal you want to use to connect the
crimp terminal and cable.

• Use crimp tools that are appropriate for the

diameter of the cable to be connected.

Advertising