9) menu k: diagnostic function setting items, 11) menu n: loop test setting items – Yokogawa ADMAG AXF User Manual

Page 62

Advertising
background image

IM 01E20C01-01E

6-33

6. PARAMETER DESCRIPTION

(9) Menu K: Diagnostic Function Setting items

Menu K contains items that are relevant to the diagno-
sis of insulation adhesion to the electrode.

[K10: Adhesion Check] Setting of adhesion
diagnostic function
This parameter selects whether or not the adhesion
diagnostic function will be carried out.

Setting

Function

Halt the adhesion diagnostic function

No

Yes

T0642.EPS

Carry out the adhesion diagnostic function

NOTE

Adhesion Diagnostic Function

᭿ This function diagnose adhesion using elec-

trode resistance values.

᭿ When “Adhesion check” has been set for B41/

H11: Display Select 2 or B42/H12: Display
Select 3
, the diagnose adhesion is indicated
on the display unit using four different levels.

᭿ If the judgment value for Level 3 is exceeded,

a warning is displayed; and if the value for
Level 4 is exceeded, an alarm is displayed.

᭿ Available conductivity for this function is

limited to:
Nominal size 10 mm or smaller: 30

␮S/cm

Nominal size 15 mm or larger: 10

␮S/cm

Make sure to use the adhesion diagnostic
function with the greater conductivity than the
above mentioned value.

F0619.EPS

Alarm

Level: 4

Warning

Level: 3

Level: 2

Level: 1

[K11: Adhesion Level1] Setting the resistance value
for adhesion level 1.
This parameter sets the resistance value (in M ohm) for
judgment of Level 1.

[K12: Adhesion Level2] Setting the resistnce value
for adhesion diagnostic level 2.
This parameter sets the resistance value (in M ohm) for
judgment of Level 2.

[K13: Adhesion Level3] Setting the resistance value
for adhesion diagnostic level 3.
This parameter sets the resistance value (in M ohm) for
judgment of Level 3.
*: The warning 80: Adhesion Wng is displayed

when the adhension level reaches Level 3.

*: If “Warning Output” has been selected for F10:

SO1 Function or F11: SO2 Function, then
status output will be performed when the adhesion
level reaches Level 3.

[K14: Adhesion Level4] Setting the resistance value
for adhesion diagnostic level 4.
This parameter sets the resistance value (in M

ohm) for

judgment of Level 4.
*: The process alarm 33: Adhesion Alm is displayed

when the adhension level reaches Level 4.

*: Alarm output will be performed if “Yes” has been

selected for G34: Alm-Adhesion.

[K15: Adh Measure Value] Displays the resistance
value for adhesion diagnose
This parameter displays the value measured using the
adhesion diagnostic function (in M

ohm).

When “No” is selected for K10: Adhesion Check,
this parameter displays the indetermination value.

(10) Menu M: Automatic Zero Adjustment

Function Setting items

Menu M contains items that are relevant to automatic
adjustment of the zero point.

[M10: Auto Zero Exe] Execution of automatic zero
adjustment function
→ Refer to the description of parameter B50.

[M11: Magflow Zero] Results of automatic zero
adjustment
This parameter is used to display the results obtained
from B50/M10: Auto Zero Exe. Specifically, the
correction value displayed, and it is also possible to
directly enter correction value.

(11) Menu N: Loop Test Setting items

Menu N contains items that are relevant to loop testing.

[N10: Test Mode] Setting for loop test execution

Setting

Function

No execution of loop testing

Loop testing is started

Normal

Test

T0644.EPS

Advertising