6) menu g: alarm setting items, 6) menu g: alarm setting items -25 – Yokogawa ADMAG AXF User Manual

Page 54

Advertising
background image

IM 01E20C01-01E

6-25

6. PARAMETER DESCRIPTION

Parameter setting sequence (for multiple ranges switching via external status input)

F10: SO1 Function
Select a function

F11: SO2 Function
Select a function

F14: SO1/2 Active Mode
Select whether SO1 and SO2 output is to be “Closed (on) Act” or
“Open (off) Act.”

Function Selection
No Function: Output is stopped (*: other functions may be selected)
Fwd/Rev Ranges: Forward and reverse flow rate measurement
Ext 2 Answer: Answerback; 2-ranges switching via external status input
Ext 3 Answer: Answerback; 3-ranges switching via external status input
Ext 4 Answer: Answerback; 4-ranges switching via external status input

Function Selection
No Function: No input function (*: other functions may be
selected)
Ext 2 Ranges: 2-ranges switching via external status input
Ext 3 Ranges: 3-ranges switching via external status input
Ext 4 Ranges: 4-ranges switching via external status input

Refer to Table 6.4.2 and set F10
and F11 or F12 and F13.

Set the spans for the ranges to be used.
No. 1 range

Յ No. 2 range Յ No. 3 range Յ No. 4 range

F12: SI1 Function
Select a function

F13: SI2 Function
Select a function

F15: SI1/2 Active Mode
Select whether SI1 and SI2 input is to be “Short Active”
or “Open Active.”

F33:Reverse

Span1

Reverse No. 1 range

B23:Flow Span

Forward No. 1 range

F0609.EPS

F34:Reverse

Span2

Reverse No. 2 range

F35:Reverse

Span3

Reverse No. 3 range

F36:Reverse

Span4

Reverse No. 4 range

F30:Forward

Span2

Forward No. 2 range

F31:Forward

Span3

Forward No. 3 range

F32:Forward

Span4

Forward No. 4 range

If “instantaneous flow rate % display (FR)” is selected for the Display
Mode and A10: FLOW RATE(%) is selected for BRAIN communication,
the instantaneous flow rate % and following symbols will be displayed
only for multiple ranges and forward/reverse flow measurement.

Forward No. 1 range : [F1]
Forward No. 2 range : [F2]
Forward No. 3 range : [F3]
Forward No. 4 range : [F4]

Reverse No. 1 range : [R1]
Reverse No. 2 range : [R2]
Reverse No. 3 range : [R3]
Reverse No. 4 range : [R4]

[F40: Auto Range Hys] Setting of automatic range-
switching hysteresis width
Automatic switching takes place for multiple ranges
switching when 100% of the range is exceeded, and
this parameter allows a hysteresis width to be set for
this switching.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.

[F41: Bi Direction Hys] Setting of forward/reverse
flow measurement hysteresis width
This parameter sets the hysteresis for forward/reverse
flow rate measurement as a % value of the minimum
flow span.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.

(6) Menu G: Alarm Setting items

(Refer to Section 6.5: Alarm Functions for more

details.)

Menu G principally contains setting items relevant to
alarms.

[G10: Low Alarm] Low alarm setting
This parameter sets the low limit (L) alarm value, and
this is done using a % value of the maximum span.
᭹ A setting value of -110% indicates that the alarm is

disabled.

[G11: High Alarm] High alarm setting
This parameter sets the high limit (H) alarm value, and
this is done using a % value of the maximum span.
᭹ A setting value of 110% indicates that the alarm is

disabled.

[G12: Low Low Alarm] Low-low alarm setting
This parameter sets the low-low limit (LL) alarm
value, and this is done using a % value of the maxi-
mum span.
᭹ A setting value of -110% indicates that the alarm is

disabled.