IM 01E20C01-01E

6-22

6. PARAMETER DESCRIPTION

[F15: SI1/2 Active Mode] Setting of the active mode
for status input
This parameter sets the active mode for the terminals
SI1 and SI2. Active modes cannot be set individually
for these two terminals.

Setting

Function

Short Active

Open Active

When the status input is set to "Short", occurrence

of the selected event will be recognized.

When the status input is set to "Open", occurrence

of the selected event will be recognized.

T0628.EPS

Example: When the “0% signal lock” function is

selected for SI1 or SI2

Signal lock status

Short Active

Signal lock status

Open Active

F0606.EPS

Short

Open

NOTE

Multiple ranges setting

Parameters from F30 to F41 are used with the
automatic multiple ranges and the multiple
ranges switching via external status input. The
followings will describe the setting method for
each range.

The multiple ranges use the following parameters:

[B23: Flow Span ] Setting of the flow rate span
(Setting of the forward No.1 range)

[F30: Forward Span 2] Setting of the forward No. 2
range

[F31: Forward Span 3] Setting of the forward No. 3
range

[F32: Forward Span 4] Setting of the forward No. 4
range

[F33: Reverse Span 1] Setting of the reverse No. 1
range

[F34: Reverse Span 2] Setting of the reverse No. 2
range

[F35: Reverse Span 3] Setting of the reverse No. 3
range

[F36: Reverse Span 4] Setting of the reverse No. 4
range

[F40: Auto Range Hys] Setting of the automatic
multiple range hysteresis width

[F41: Bi Direction Hys] Setting of the forward/
reverse flow measurement hysteresis width

Multiple Ranges Setting 1:

Automatic multiple ranges switching

᭿ When the instantaneous flow rate exceeds 100% of

the range, transition to the next range (up to four
ranges) is carried out automatically. Furthermore,
when the flow is in reverse, the reverse range can
also be automatically selected.

᭿ Range switching can be confirmed according to the

status of the SO1 and SO2 status output terminals.
Refer to Table 6.4.1: Status Output for Automatic
Multiple Ranges Switching for details of status
output conditions for each range.

Status Output for Automatic Multiple Ranges Switching

Operations are performed in accordance with the
following table when the active mode has been set to
“Closed (On) Act” using F14: SO1/2 Active Mode.
Operating patterns are reversed when the active mode
has been set to “Open (Off) Act”.

Table 6.4.1 Status Output for Automatic Multiple Ranges

Switching

Parameter setting

Function

F10

Status output

SO2

SO1

F11

T0629.EPS

Open

Closed

Open

Closed

Open

Open

Closed

Open

Open

Closed

Closed

Open

Closed

Open

Closed

Open

Open

Closed

Closed

Open

Closed

Open

Closed

Open

Closed

Open

Open

Closed

Open

Closed

Open

Open

Closed

Closed

Open

Closed

Open

Closed

No Function

Fwd/Rev
Ranges

No Function

Auto
2 Ranges

No Function

Auto
3 Ranges

Auto
4 Ranges

Fwd/Rev
Ranges

Auto
2 Ranges

No Function

No Function

No Function

Auto
2 Ranges

Auto
3 Ranges

Auto
4 Ranges

Auto
2 Ranges

Fwd/Rev
Ranges

Forward single range

Auto forward/reverse
1 range (SO1)

Auto forward/reverse
1 range (SO2)

Auto forward
2 ranges (SO1)

Auto forward

2 ranges (SO2)

Auto forward
3 ranges

Auto forward
4 ranges

Auto forward/reverse

2 ranges

Auto forward/reverse
2 ranges

*: "No Function" is the default value. Only SO1 or SO2 terminals are

used for single or dual ranges; accordingly, the unused terminal
remains at the default value (i.e., No Function) and can therefore
be used for other functions.

Forward

Reverse

Forward

Reverse

Forward 1 range

Forward 2 range

Forward 1 range

Forward 2 range

Forward 1 range

Forward 2 range

Forward 3 range

Forward 1 range

Forward 2 range

Forward 3 range

Forward 4 range

Forward 1 range

Forward 2 range

Reverse 1 range

Reverse 2 range

Forward 1 range

Forward 2 range

Reverse 1 range

Reverse 2 range

Fwd/Rev
Ranges

Advertising