IM 01E20C01-01E

6-9

6. PARAMETER DESCRIPTION

(8) Item H (Menu H): Display Setting items

Menu H contains setting items that are relevant to display on the display unit.

T0609.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

H10 Display Select1

(DISP SELECT1)

W

Flow Rate(%)
Flow Rate
Flow Rate(mA)
Forward Total
Reverse Total
Dif Total

Flow Rate

Selects content of the first line
for Display Mode. Linked with
B40.

H20 Display Cycle

(DISP CYCLE)

W

200ms
400ms
1s
2s
4s
8s

400ms

Selects the display cycle.

H30 Language

(LANGUAGE)

W

English
Japanese
French
German
Italian
Spanish

English

Selects the language used by the
display unit. Linked with B10.

H11 Display Select2

(DISP SELECT2)

W

Off
Flow Rate(%)
Flow Rate
Flow Rate(mA)
Flow Rate(Bar)
Forward Total
Reverse Total
Dif Total
Tag No
Adhesion Check
Communication

Off

Selects content of the second line
for Display Mode. Linked with
B41.

H12 Display Select3

(DISP SELECT3)

W

Same as H11
(Display Select2)

Off

Selects content of the third line
for Display Mode. Linked with
B42.

H60 —

(SELF CHECK)

R

Good
Error

See “6.5 Alarm Functions”.

H00 Display Set

(DISP SET)

(9) Item J (Menu J): Auxiliary Function Setting items

Menu J contains setting items such as the flow direction, rate limits, and low cut.

T0610-1.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

J00 Aux

(AUX)

J11 4-20mA Low Lmt

(4-20 LOW LMT)

W

-20.0 to 100.0

%

1

-20.0%

Sets the low limit for current
output

J12 4-20mA High Lmt

(4-20 HI LMT)

W

0.0 to 120.0

%

1

120.0%

Sets the high limit for current
output

J20 Flow Direction

(FLOW DIRECT)

W

Forward
Reverse

Forward

Selects the flow direction.

J21 Rate Limit

(RATE LIMIT)

W

0 to10

%

0

5%

Sets the level to reduce output
fluctuation.

J10 4-20mA Low Cut

(4-20 LOW CUT)

W

%

0 to 10

0

0%

Sets the range in vicinity of 0%
within which the current output
will be 4 mA

J23 Pulsing Flow

(PULSING FLOW)

W

No
Yes

No

Selects whether pulsing flow is to
be supported.

J24 T/P Damp Select

(T/P DAMP SEL)

W

Damping
No Damping

Damping

Selects whether the flow rate value
obtained through damping calculation
for total/pulse or the instantaneous
flow rate value (no damping) for
total/pulse is to be used.

J22 Dead Time

(DEAD TIME)

W

s

0 to 15

0

0s

Sets the dead time to reduce output
fluctuation. When “0” is set, rate
limit function is not available.