IM 01E20C01-01E

3-1

3. INSTALLATION

3.

INSTALLATION

WARNING

Installation of the magnetic flowmeter must be performed by expert engineer or skilled personnel. No
operator shall be permitted to perform procedures relating to installation.

3.1 Installation Location

IMPORTANT

Install the instrument in a location where it is not exposed to direct sunlight. For ambient temperature,
refer to Chapter 11 “OUTLINE”.
The instrument may be used in an ambient humidity where the RH ranges from 0 to 100%. However,
avoid long-term continuous operation at relative humidity above 95%.

3.2 Mounting

This instrument can be mounted using surface mounting, 2-inch pipe mounting, or panel
mounting.

Surface Mounting (Wall Mounting)

4-

␾6 Hole or

M6 Screw

194 (7.64)

65

(2.56)

F0301.EPS

For surface mounting, use the mounting
fixture provided, using M6 screws.
These M6 screws must be provided
by the user.

Unit: mm

(approx. inch)

Figure 3.2.1 Surface Mounting

IMPORTANT

Mounting fixture on equipment intended to be mounted on a wall or ceiling shall withstand a force of four
times the weight of the equipment (AXFA11: 3.3kg (7.3 lb)).