Normal operating conditions, Caution – Yokogawa ADMAG AXF User Manual

Page 105

Advertising
background image

IM 01E20C01-01E

11-4

11. OUTLINE

Size 2200 mm (88 in.) to 2600 mm (104 in.)

2.4

2.0

1.6

1.2

0.85

0.8
0.4

0

0

1

2

6

8

10

V[m/s]

F16-3.EPS

Pulse
Output
Accuracy

% of Rate

High grade Accuracy

High grade Accuracy

Standard Accuracy

Current Output: Pulse output accuracy plus

Ϯ0.05% of Span

Repeatability:

Ϯ0.1% of Rate (V м 1 m/s (3.3 ft/s))

Ϯ0.05% of Rate Ϯ0.5 mm/s (V < 1 m/s (3.3 ft/s))

Maximum Power Consumption:

Combined with AXF Remote Flowtube: 20 W

Insulation Resistance(*1) :

Between power supply terminals and ground
terminal: 100M

Ω at 500V DC

Between power supply terminals and input/output/
excitation current terminals : 100M

Ω at 500V DC

Between ground terminal and input/output/excitation
current terminals: 20M

Ω at 100V DC

Between input/output/excitation current terminal: 20M

at 100V DC

Withstand Voltage(*1) :

Between power supply terminals and ground terminal:
1390V AC for 2 seconds
Between power supply terminals and input/output
terminals: 1390V AC for 2 seconds
Between excitation current terminal and ground
terminal: 160V AC for 2 seconds
Between excitation current terminal and input /output
terminals: 200V AC for 2 seconds

CAUTION

*1: When performing the Insulation Resistance Test or the

Withstand Voltage Test, please obey the following
caution.
• Following the relevant test, wait for more than 10

seconds after the power supply has been turned off
before removing the cover.

• Remove all wires from terminals before testing.

• When the power terminal has a lighting protector

(optional code A), remove the short bar at the ground
terminal.

• After testing, be sure to discharge by using a resistance

and return all wires and the short bar to its correct
position.

• Screws must be tightened to a torque of 1.18 N-m or

more.

• After closing the cover, the power supply can be

restored.

Safety Requirement Standards:

EN61010-1

• Altitude at installation site: Max. 2000 m above sea level
• Installation category based on IEC1010:

Overvoltage category II (“II” applies to electrical equip-
ment which is supplied from a fixed installation-like
distribution board.)

• Pollution degree based on IEC1010

Pollution degree 2 (“Pollution degree” describes the
degree to which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is adhering. “2”
applies to a normal indoor atmosphere.)

EMC Conformity Standards:

,

EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
EN61000-3-2 ClassA
EN61000-3-3

CAUTION

This instrument is a Class A product, and it is
designed for use in the industrial environment.
Please use this instrument in the industrial
environment only.

NORMAL OPERATING CONDITIONS

Ambient Temperature: –40

°C to +60°C (–40°F to +140°F)

Indicator’s operating range: –20

°C to +60°C (–4°F to +140°F)

Ambient Humidity: 0 to 100%

Lengthy continuous operation at 95% or more is not
recommended.

Power Supply:

Power supply code 1:

• AC specifications

Rated power supply: 100 to 240 V AC, 50/60 Hz
(Operating voltage range: 80 to 264 V AC)

• DC specifications

Rated power supply: 100 to 120 V DC
(Operating voltage range: 90 to 130 V DC)

Power supply code 2:

• AC specifications

Rated power supply: 24 V AC, 50/60 Hz
(Operating voltage range: 20.4 to 28.8 V AC)

• DC specifications

Rated power supply: 24 V DC
(Operating voltage range: 20.4 to 28.8 V DC)

Supply Voltage and

Cable Length for Power Supply Code 2

0

20.4

22

24

26

28.8

100 ( 330)

200 ( 660)

280 ( 920)

420 (1380)

F00.EPS

Cable cross section area: 1.25 mm

2

Cable cross section area: 2 mm

2

Allo

w

ab

le cab

le length m(ft)

Usable range E (V)