Alarms & Warning Messages

System Alarms (Device breakdown or inability to obtain correct measurements.)

Display unit/BRAIN
(

ᮀ60) content

Alarm countermeasure
message on display unit

10:uP Fault

Contact nearest office or
service center

Microprocessor (CPU) failure

11:EEPROM Fault

Contact nearest office or
service center

EEPROM failure

12:A/D(H) Fault

Contact nearest office or
service center

A/D converter failure

13:A/D(L) Fault

Contact nearest office or
service center

14:A/D(Z) Fault

Contact nearest office or
service center

15:Coil Open

Cut the power and check coil
& EX cable

Flowtube coil is open-circuit

16:EEPROM Dflt

Contact nearest office or
service center

EEPROM default values

Contact your nearest Yokogawa office or
service center.

Turn off the AXFA11 power supply and
examine the excitation cable for breakage.
If there is no breakage, contact your nearest
Yokogawa office or service center.

Contact your nearest Yokogawa office or
service center.

Alarm description

Countermeasure

T0651.EPS

Process Alarms (Device is normal but process-related errors make correct measurement impossible.)

Display unit/BRAIN
(

ᮀ60) content

Alarm countermeasure
message on display unit

30:Sig Overflow

Check signal cable
and grounding

Input signal error

31:Empty Pipe

Fill flow tube with
fluid

Flowtube is not filled with
fluid

Carry out an investigation as follows:

• Check the signal cable for breakage.

• Check for contact between signal cable,

power cable, and excitation cable.

• Check for stray currents in the fluid.

• Check the grounding.

Fill the flowtube with fluid.

33:Adhesion Alm

Clean electrodes

Electrode adhesion alarm

Clean the electrodes.

32:HH/LL Alm

Check the flow rate and
setting value

Flow rate alarm for greater
than High-High limit value
or less than Low-Low limit
value.

Check the flow rate and setting value of
High High limit and Low Low limit.

Alarm description

Countermeasure