Yokogawa ADMAG AXF User Manual
Page 96

IM 01E20C01-01E
9-4
9. ACTUAL OPERATION
Touch the
switch twice to move the cursor
to “Ext Auto Zero”.
Touch the
switch to select “Ext Auto Zero
(Zero adjustment via external status input)”.
In order to request confirmation, the entire display
flashes on and off. Touch the
switch once
again at this time to fix selection of the automatic
zero adjustment function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except
are carried
out, the parameter cannot be set.
The system automatically returns to sub-item
selection screen (E).
Zero adjustment will be started if the SI1 terminals are shorted (When the active mode has
been set to “Short Active” using F15: SI1/2 Active Mode).
This process will end after approximately 30 seconds.
Auto zero adjustment function is being executed.
(about 30 seconds.)
NOTE
When the SI1 terminals continue to be shorted, this zero adjustment is automati-
cally repeated.
F0902-2.EPS
F:Status Function
11:SO2 Function
12:SI1 Function
13:SI2 Function
F12:SI1 Function
Ext Auto Zero
Sub-item
Parameter
Search Mode
F12:SI1 Function
No Function
Ext Auto Zero
Ext Ttl Preset
Major item
Sub-item
selection (E)
F12:SI1 Function
No Function
No Function
0% Signal Lock
X2
Parameter
Replacement
(execute) Mode
F0902-3.EPS
Now Auto Zero