3 wiring ports, Wiring ports -3 – Yokogawa ADMAG AXF User Manual
Page 13

IM 01E20C01-01E
4-3
4. WIRING
4.3 Wiring Ports
This instrument is of watertight construction as
stipulated in JIS C0920. It is shipped with a wiring
bracket (waterproof gland or waterproof gland with
union) or a plastic gland attached, only in cases where
an optional specification is selected for the wiring port.
IMPORTANT
The wiring port is sealed with a cap (not water-
proof). Do not remove the cap from the unused
wiring port. If waterproof property is necessary,
please use waterproof glands.
(1) When waterproof property is unnecessary
(When there are no particular optional
specifications)
The wiring port is sealed with a cap (not water-proof)
that must be removed before wiring. At this time,
handle the wiring port in accordance with the JIS
C0920 mentioned above. Do not remove the cap from
the unused wiring port.
(2) When waterproof property is necessary
(Wiring using waterproof glands)
IMPORTANT
To prevent water or condensation from entering
the converter housing, waterproof glands are
recommended. Do not over-tighten the glands or
damage to the cables may result. Tightness of
the gland can be checked by confirming that the
cable is held firmly in place.
Cable
Waterproof gland
Gasket
Washer
F0404.EPS
Figure 4.3.1 Waterproof Gland (Optional Code EG)
For working on the electric wire tubes or the flexible
tubes (PF1/2), remove the waterproof gland and attach
them directly to the wiring port.
Gasket
Washer
Tightening gland
When working on conduit pipes or flexible pipes (PF1/2 only)
Cable
PF1/2
F0405.EPS
Figure 4.3.2 Waterproof Gland with Union Joint
(Optional Code EU)
Plastic gland
F0406.EPS
Figure 4.3.3 Plastic Gland (Optional Code EP)
Gasket
Washer
When working on electric wire tube or flexible tube (PF3/4)
* When connecting PF1/2, remove the conversion plug and
connect directly to wiring port.
Extension plug (x2)
Cable
PF3/4
PF1/2
Conversion plug (x5)
Figure 4.3.4 PF3/4 Waterproof Gland (Optional Code EW)