Yokogawa ADMAG AXF User Manual

Page 40

Advertising
background image

IM 01E20C01-01E

6-11

6. PARAMETER DESCRIPTION

(12) Item N (Menu N): Loop Test Setting items

Menu N contains items that are relevant to the execution of loop testing.

T0613.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

N10 Test Mode

(TEST MODE)

N00 Test

(TEST)

W

Normal
Test

Normal

Selects whether mode will be set
to “Normal” or “Test”.

N20 Test SO1

(TEST SO1)

W

Open(Off)
Closed(On)

Open(Off)

Selects the test condition for
SO1 terminal.

N21 Test SO2

(TEST SO2)

W

Open(Off)
Closed(On)

Open(Off)

Selects the test condition for
SO2 terminal.

N22 Test Alarm Out

(TEST ALM OUT)

W

Open(Off)
Closed(On)

Closed(On)

Selects the test condition for
alarm output terminal.

N23 Test SI1

(TEST SI1)

R

Open
Short

Displays the test condition for
SI1 terminal.

N24 Test SI2

(TEST SI2)

R

Open
Short

Displays the test condition for
SI2 terminal.

N60 —

(SELF CHECK)

R

Good
Error

See “6.5 Alarm Functions”.

N11 Test Output Value

(TEST OUT VAL)

W

%

-10 to 110

0

0%

Sets the test output value.

(13) Item P (Menu P): Parameter Protection items

Menu P contains items that are relevant to write protection and passwords.

T0613-1.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

P10 Key Code

(KEY CODE)

P00 Protect

(PROTECT)

W

0 to 9999

0

Parameter of the display restriction

P20 Write Protect

(W PROTECT)

R

No
Yes

No

Displays whether or not overwriting
of parameter data is prohibited.

P21 Enable Wrt Passwd

(ENABLE WRITE)

W

ASCII 8 characters

Sets the correct password so that
write protection function will be
released.

P22 New Password

(NEW PASSWORD)

W

ASCII 8 characters

Sets the password for write protection
funcion

P23 Software Seal

(SOFT SEAL)

R

Break
Keep

Displays whether or not a Joker
password was used (Break).

P60 —

(SELF CHECK)

R

Good
Error

See “6.5 Alarm Functions”.

Keep

Advertising