Yokogawa ADMAG AXF User Manual

Page 52

Advertising
background image

IM 01E20C01-01E

6-23

6. PARAMETER DESCRIPTION

Parameter setting sequence

(for automatic multiple ranges switching)

F10: SO1 Function
Select a function

F11: SO2 Function
Select a function

F14: SO1/2 Active Mode
Select whether SO1 and SO2 output is to
be “Closed (On) Act” or “Open (Off) Act”

Function Selection
No Function: Output is stopped (*: other functions may be selected)
Fwd/Rev Ranges: Forward and reverse flow rate measurement
Auto 2 Ranges: Automatic 2-ranges switching
Auto 3 Ranges: Automatic 3-ranges switching
Auto 4 Ranges: Automatic 4-ranges switching

Refer to Table 6.4.1
and set F10 or F11.

Set the spans for the ranges to be used.
No. 1 range

Ϲ No. 2 range

Ϲ No. 3 range Ϲ No. 4 range

If “instantaneous flow rate % display (FR)” is selected for
Display Mode and A10: FLOW RATE(%) is selected for
BRAIN communication, the instantaneous flow rate %
and following symbols will be displayed only for multiple
ranges and forward/reverse flow measurement.
Forward No. 1 range : [F1] Reverse No. 1 range : [R1]
Forward No. 2 range : [F2] Reverse No. 2 range : [R2]
Forward No. 3 range : [F3]
Forward No. 4 range : [F4]

F30: Forward

Span 2

Forward No. 2 range

B23: Flow Span

Forward No. 1 range

F31: Forward

Span 3

Forward No. 3 range

F33: Reverse

Span 1

Reverse No. 1 range

F34: Reverse

Span 2

Reverse No. 2 range

F32: Forward

Span 4

Forward No. 4 range

F0607.EPS

10% of No. 1 range
(set using F40)

10% of No. 2 range
(set using F40)

10% of No. 3 range
(set using F40)

Input

100%

No. 1 range No. 2 range No. 3 range No. 4 range

Output

Reverse

Forward

Minimum span % within multiple and
forward/reverse (set using F41)

F0608.EPS

Figure 6.4.1 Multiple Ranges and Hysteresis Widths

NOTE

For more details regarding the setting of hyster-
esis width, refer to the description of setting
parameter for F40: Auto Range Hys and F41:
Bi Direction Hys
.

Multiple Ranges Setting 2:

Multiple ranges switching via external
status input

᭿ For both the forward and reverse directions, switch-

ing of up to four ranges can be carried out based on
status input; however, switching between directions
is not possible. Switching between forward and
reverse ranges is carried out automatically only
when the flow direction reverses.

᭿ SI1 and SI2 status input terminals are used for

multiple range switching. For more details, refer to
Table 6.4.2: Multiple Ranges Switching via External
Status Input.