<2. Handling Cautions>

2-2

IM 01C27B01-01EN

NOTE

When storing the instrument with a battery

pack, it is recommended to put the instrument in

Deep Sleep mode to conserve the batteries. For

details on how to switch to Deep Sleep mode,

refer to subsection 8.3.16 “Switching to Deep

Sleep Mode”.

2.4 Selecting the Installation

Location

The transmitter is designed to withstand severe

environmental conditions. However, to ensure

that it will provide years of stable and accurate

performance, take the following precautions when

selecting the installation location.
(a) Wireless Communication

NOTE

The installation location of this transmitter must

meet the following conditions:

- Adjust the direction of the antenna to be

in the upright position regardless of the

orientation of this transmitter. See section 4

for adjusting the antenna.

- Install the transmitter at least 1.5m above

the ground or floor.

F0203.ai

1.5m or more

- Ensure that there are no obstacles such as

walls or pipes within a 30-cm radius of each

antenna.

- Confirm that each field wireless equipment

compliant with ISA100.11a can see the

antenna of other devices which locate within

its own communication range. In the star

topology network, the visibility to the antenna

of gateway is a mandatory clause.

(b) Ambient Temperature

Avoid locations subject to wide temperature

variations or a significant temperature gradient.

If the location is exposed to radiant heat from

plant equipment, provide adequate thermal

insulation and/or ventilation.

(c) Ambient Atmosphere

Do not install the transmitter in a corrosive

atmosphere. If this cannot be avoided, there

must be adequate ventilation.

(d) Shock and Vibration

Although the transmitter is designed to be

relatively resistant to shock and vibration, an

installation site should be selected where this is

kept to a minimum.

(e) Installation of Explosion-protected Transmitters

An explosion-protected transmitters is

certified for installation in a hazardous area

containing specific gas types. See subsection

2.8 “Installation of an Explosion-Protected

Transmitters.”

2.5 Pressure Connection

WARNING

• Never loosen the process connector bolts

when an instrument is installed in a process.

The device is under pressure, and a loss of

seal can result in a sudden and uncontrolled

release of process fluid.

• When draining toxic process fluids that have

condensed inside the pressure detector,

take appropriate steps to prevent the contact

of such fluids with the skin or eyes and the

inhalation of vapors from these fluids.