Yokogawa EJX430B User Manual

Page 78

Advertising
background image

<9. Maintenance>

9-13

IM 01C27B01-01EN

Integral

indicator

Factory

NAMUR

category

Bit

Diagnostic

Status

Diagnostic Status Detail

Cause

Release/

recovery

conditions

(except

restart)

Action

AL. 63

AI SIM

C

Bit 17 Simulation is

active

SimulationActive

(AI1)

AI1 block is

simulate mode.

Returns when

the simulate

mode of AI1

block is set

to disable. 1

(Disable)

Check the

simulate

mode of AI1

block.

AL. 64

AI SIM

SimulationActive

(AI2)

AI2 block is

simulate mode.

Returns when

the simulate

mode of AI2

block is set

to disable. 1

(Disable)

Check the

simulate

mode of AI2

block.

AL. 65

AI SIM

C

Bit 17 Simulation is

active

SimulationActive

(AI3)

AI3 block is

simulate mode.

Returns when

the simulate

mode of AI3

block is set

to disable. 1

(Disable)

Check the

simulate

mode of AI3

block.

*1:

"Factory NAMUR category” refers to the four categories (C: Check function, M: Maintenance required, F: Failure, and O: Off

specification) according to NAMUR NE107*.

* NAMUR NE107 “Self-Monitoring and Diagnosis of Field Devices”

*2: When the device detects “AL01 CAP.ERR” and “AL02 AMP.ERR”, the LCD display stays on regardless of the status in LCD mode.

*3: Applicable for Integral antenna type (Amplifier housing code: 7).

*4: Applicable for Detachable antenna type (Amplifier housing code: 8 or 9).

Advertising