<8. Setting Parameters>

8-1

IM 01C27B01-01EN

8. Setting Parameters

This transmitter can remotely handle range

changes, Tag No. setup, monitoring of self-

diagnostic results, and Zero Point Adjustment, etc.

according to communication with the field wireless

configuration tool or the device configuration tool.

8.1 Environment for parameter

setting

After installing the battery pack, perform

provisioning and have the transmitter join the field

wireless network.

This transmitter supports the OOB (out-of-band)

method using the infrared communication as a

provisioning method.

Refer to Subsection 7.4 “Connecting to the Field

Wireless Network” for details of the provisioning.

F0801.ai

Infrared port

Figure 8.1

Infrared port of the transmitter

8.2 Preparing Software

8.2.1 Softwares for the Field Wireless

Configuration Tool and the Device

Configuration Tool

Before using the device configuration tool,

confirm that CF/DD and DeviceDTM for your

transmitter (wireless EJX) are installed in the device

configuration tool.

CF/DD and DeviceDTM are provided by DVD-ROM

attached to YFGW710 Field Wireless Integrated

Gateway.

Refer to YFGW710 Field Wireless Integrated

Gateway (IM 01W01F01-01EN) for details.

CF (Capabilities File)/DD (Device Description)

A CF file contains information, such as the

vendor of the field device, its model and

revision, available types of process data (flow

rate, temperature, pressure, etc.), and number

of data items. A DD file contains the information

on parameters, such as data structures and

attributes.

DeviceDTM

DeviceDTM (Device Type Manager) is driver

software for field devices provided based on the

FDT (Field Device Tool) technology.

The field wireless configuration tool or the device

configuration tool allows confirming the device

information.

Refer to Subsection 9.2 “Calibration Instruments

Selection” for the field wireless configuration

tool or the device configuration tool of our

recommendation.

8.2.2 Software Download

Software download function permits to update

wireless field device software via ISA100.11a

wireless communication. For details, refer to

YFGW710 Field Wireless Integrated Gateway (IM

01W01F01-01EN) or YFGW410 Field Wireless

Management Station (IM 01W02D01-01EN).

8.3 Setting Parameters

8.3.1 Parameter Usage and Selection

Before setting a parameter, please see the following

table for a summary of how and when each

parameter is used.

IMPORTANT

After setting and sending data with the field

wireless configuration tool or the device

configuration tool, wait 30 seconds before

turning off the transmitter. If it is turned off

too soon, the settings will not be stored in the

transmitter.