background image

<9. Maintenance>

9-15

IM 01C27B01-01EN

Integral

Indicator

Factory

NAMUR

category

Bit

Diagnostic

Status

Diagnostic Status Detail

Output actions

Pressure

Static

Pressure

Capsule

Temp

Value

Amp

Temp

Value

AL. 10

PRESS

O

Bit 23 Outside sensor

limits

DP_OUTSIDE_LIMIT

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Sensor

Conversion

not Accurate)

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Non Specific)

Normal action Normal action

AL. 11

ST. PRSS

SP_OUTSIDE_LIMIT

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Non Specific)

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Sensor

Conversion

not Accurate)

Normal action Normal action

AL. 12

CAP. TMP

CAPT_OUTSIDE_LIMIT

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Non Specific)

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Non Specific)

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Sensor

Conversion

not Accurate)

Normal action

AL. 13

AMP. TMP

AMPT_OUTSIDE_LIMIT

Normal action Normal action Normal action

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Sensor

Conversion

not Accurate)

AL. 53

P. ADJ

C

Bit 25

Installation,

calibration

problem

DP_TRIM_SPAN_

OUTSIDE

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Range Limits

Exceeded)

Normal action Normal action Normal action

AL. 53

P. ADJ

DP_TRIM_ZERO_

OUTSIDE

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Range Limits

Exceeded)

Normal action Normal action Normal action

AL. 55

SP. ADJ

SP_TRIM_SPAN_

OUTSIDE

Normal action

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Range Limits

Exceeded)

Normal action Normal action

AL. 55

SP.ADJ

SP_TRIM_ZERO_

OUTSIDE

Normal action

Output value

(calculated in

a normal way)

Output status

(UNCERTAIN:

Range Limits

Exceeded)

Normal action Normal action

AL. 79

OV. DISP

LCD_OUTSIDE_LIMIT

Normal action Normal action Normal action Normal action