<8. Setting Parameters>

8-19

IM 01C27F01-01EN

The low cut point has hysterisis so that the output

around the point is behaved as below figure.

<Example>

Output mode: Linear

Low cut mode: Zero

Low cut: 20.00%

Output

Setting range:

0 to 20%

Input

0%

(20%)

2%

Low

cut

point

Hystrersis fixed at 10%
of the cut point

F0805.ai

8.3.8 Integral Indicator Scale Setup

The following two displays are available for the

integral indictor: pressure and temperature. The

following two variables can be displayed on the

integral indicator: % of pressure range, and % of

temperature range.

Available displays

Description

and related parameters

% of range

(PRES %)

Indicates input pressure in –10 to

110% range depending on the set

range (

LRV and URV).

PRES % 45.6 %

Follow the procedure described in (1) to (2) below

to set the integral indicator.

(1) Display Selection
Display set to Display Selection is displayed on the

integral indicator.

• Procedure to call up the Display Selection

display

TRANSDUCER block: Display Selection

The Display Selection parameter enables the

pressure (AI1 block), and, temperature (AI3

block) to be displayed on the LCD. Select

whether or not to enable each block to be

displayed.

(2) Cyclic Display
Information in the AI1, AI3 blocks can be displayed

cyclically according to the display On/Off setting

for the pressure (AI1 block), and temperature (AI3

block) selected in the Display Selection parameter.

8.3.9 Unit for Displayed Temperature

When the instrument is shipped, the temperature

units are set to C (Centigrade). Follow the

procedure below to change this setting.

• Procedure to call up the Sensor Range.Units

Index display

AI3 Block: Sensor Range.Units Index

Confirm that ºC (deg C) is selected in the Units

Index parameter for the temperature(AI3 block).

Note: When the unit is changed by Sensor Range.Units Index

parameter in temperature (AI3 block), units of capsule

temperature is also changed.

8.3.10 Zero Point Adjustment and Span

Adjustment

Each EJX-B series transmitter is characterized

by factory. But there are some errors caused by

environment and installed posture.
There are Zero and Span Adjustments to fine-tune

those errors.

Zero Adjustment is adjustment for one point to

adjust the bottom value of the measurement range

as 0% of output.
The Span Adjustment defines input and output

characteristic between two points that’s one side

assumed as standard.

This is used when there is doubt of span drift or

when it is impossible to make zero at absolute

pressure with adjustment for user’s pressure

standard.