<10. Parameter Summary>

10-1

IM 01C27F01-01EN

10. Parameter Summary

The setting of differential pressure (Al2 block) is invalid in this instrument. The output value of Al2 object and

the output status are “0.0”(zero) and “Uncertain” respectively.

Various diagnostics of the Al2 object and the static pressure is not supported in this instrument.

Table 10.1

Parameter

Object

ID

Attribute

ID

Label

Description

Default

value

Handling

1.

UAPMO

block

1

Version Revision

Indicates the application revision of EJX This revision

when the application software is downloaded.

---

R

10

Static Revision

Indicates the revision level of the fixed parameters of

UAP Used, for example, to check whether parameters

have been change.

0

R

64

Identification

Number

Indicates the vender ID, model ID, and revision of the

device.

---

R

65

CTS Version

Indicates the version of the communication stack test

system (CTS).

0

R

66

ITS Version

Indicates the version of the inter operability test system

(ITS).

0

R

67

Diagnostic Status

Indicates the diagnostic results of the device based on

the NAMUR NE107

*1

model.

Setting Enable diagnostic status configuration in UAP

Option to Enable allows turning OFF and ON the

display of the diagnostic results for each summary,

and changing Categorize For Categorize at the time of

shipment, refer to tables 8.3 and 10.2.

---

R

68

UAP Option

Allows setting the Diagnostic Status and write protection

of UAP.

Software write protect

1: On, 0: Off

Enable hardware write protect

1: Enable, 0: Disable (default)

Enable diagnostic status configuration

1: Enable, 0: Disable (default)

The following table shows the relationship between the

hardware write protection and software write protection.

Enable

hardware

write

protect

Hardware

write

protect

Software

write

protect

Write

protect

Disable

Off or On

Off

No

(Writable)

Disable

Off or On

On

Yes

(Unwritable)

Enable

Off

Off or On

No

(Writable)

Enable

On

Off or On

Yes

(Unwritable)

0

W

69

Diagnostic

Switch

Allows setting On/Off for each summary of Diagnostic

Status when Enable diagnostic status configuration in

UAP Option is set to Enable.

On

W

70

Diagnostic

Configuration

Allows Categorize for each summary of Diagnostic

Status when Enable diagnostic status configuration in

UAP Option is set to Enable.

Refer to Table

10.2.

W

102

Diagnostic Status

Detail[2]

Detailed information on Diagnostic Status. For

Categorize at the time of shipment, refer to Table 10.2.

Refer to Table

10.2.

R

103

Energy Left

Indicates the number of days of remaining battery life

assuming ambient temperature condition as 23 degrees

Celsius.

The unit is day.

---

R