<10. Parameter Summary>

10-4

IM 01C27F01-01EN

Object

ID

Attribute

ID

Label

Description

Default

value

Handling

3.

CO

block

(continued)

3

COMM_

CONTRACT

Indicates the Contract information. The following

shows the components.

1. ContractID

2. Contract_Status

3. Actual_Phase

---

R

4

PUB_ITEM_MAX

Maximum PUB_ITEM value

---

R

5

PUB_ITEM_NUM

PUB_ITEM number

---

R

6

PUB_ITEM

Indicates the PUB_ITEM information. The following

shows the components

1. ObjectID

2. AttributeID

3. AttributeIndex

4. Size

---

W

4.

TRANSDUCER

block

1

Tag Description

Memo field available to write anything.

Transducer

W

2

Auto Recovery

Allows specifying the action when the cause of a

sensor failure is removed.

OFF (=0):

AL continues to be indicated

even after the cause of the sensor

error is removed, and the BURN

OUT state is not released.

ON (=1):

The AL indication for sensor

failure disappears and normal

action is resumed when the cause

of the sensor failure is removed.

ON

W

3

Model

Indicates the model name of the transmitter.

---

R

4

Sensor Serial

Number

Indicates the device number of the transmitter.

---

R

5

Measurement

Rate

*3

Indicates the measurement period of process values.

---

R

6

Measurement

Mode

Measurement mode selects continuous mode or

intermittent mode.

When the update time is set to 0.5 second

*4

in

intermittent mode, the mode is automatically switched

to continuous mode.

When the update time is set to 0.5 second

*4

,

the computation process is in continuous mode,

regardless of the measurement mode.

When 0 is set, measurement period is set to 30

seconds regardless of measurement setting.

intermittent

mode

W

7

Wireless Status

Indicate the communication status.

1. Indicates the initial idle status or join status.

(idle status, join status)

2. Indicates whether Contract(Pub) is established.

(not establishment, establishment)

3. Indicates whether Contract(R/W) is

established.

(not establishment, establishment)

1. idle status

2. not

estabilishment

3. not

esabilishment

R

8

Display Selection

Select PV Value displaying on the LCD indicator.

1. Indicates display status of PV Value of AI1

(displayed, not displayed)

2. Indicates display status of PV Value of AI2

(displayed, not displayed)

3. Indicates display status of PV Value of AI3

(displayed, not displayed)

1. displayed

2. not

displayed

3. not

displayed

W

9

LCD Mode

Select the display mode on the LCD indicator.

1. Indicates the ON mode. (off, intermittent)

2. Indicates whether continuous or not.

(continuous off, continuous on)

*3

3. Indicates the bar graph display.

(bar graph not displayed, bar graph displayed)

1. OFF

2. continuous

OFF

*3

3. bar graph

not displayed