1 safe use of this product, Safe use of this product -2 – Yokogawa EJX530B User Manual

Page 6

Advertising
background image

<1. Introduction>

1-2

IM 01C27F01-01EN

Functional grounding terminal

Caution

This symbol indicates that the operator

must refer to an explanation in the user’s

manual in order to avoid the risk of injury

or death of personnel or damage to the

instrument.

1.1 Safe Use of This Product

For the safety of the operator and to protect the

instrument and the system, please be sure to follow

this manual’s safety instructions when handling this

instrument. If these instructions are not heeded,

the protection provided by this instrument may be

impaired. In this case, Yokogawa cannot guarantee

that the instrument can be safely operated. Please

pay special attention to the following points:

(a) Installation

• This instrument may only be installed by an

engineer or technician who has an expert

knowledge of this device. Operators are not

allowed to carry out installation unless they

meet this condition.

• With high process temperatures, care must

be taken not to burn yourself by touching the

instrument or its casing.

• Never loosen the process connector nuts when

the instrument is installed in a process. This can

lead to a sudden, explosive release of process

fluids.

• When draining condensate from the pressure

detector section, take appropriate precautions

to prevent the inhalation of harmful vapors and

the contact of toxic process fluids with the skin

or eyes.

• When removing the instrument from a

hazardous process, avoid contact with the fluid

and the interior of the meter.

• All installation shall comply with local installation

requirements and the local electrical code.

(b) Wiring

• The instrument must be installed by an

engineer or technician who has an expert

knowledge of this instrument. Operators are not

permitted to carry out wiring unless they meet

this condition.

(c) Maintenance

• Please carry out only the maintenance

procedures described in this manual. If you

require further assistance, please contact the

nearest Yokogawa office.

• Care should be taken to prevent the build up of

dust or other materials on the display glass and

the name plate. To clean these surfaces, use a

soft, dry cloth.

(d) Explosion Protected Type Instrument

• Users of explosion proof instruments should

refer first to Section 2.8 (Installation of an

Explosion-Protected Instrument) of this manual.

• The use of this instrument is restricted to those

who have received appropriate training in the

device.

• Take care not to create sparks when accessing

the instrument or peripheral devices in a

hazardous location.

• Repair or modification to this instrument by

customer will cause malfunction of explosion

protect function and hazardous situation. If you

need to repair or modification, please contact

the nearest Yokogawa office.

(e) Modification

• Yokogawa will not be liable for malfunctions or

damage resulting from any modification made

to this instrument by the customer.