<9. Maintenance>

9-5

IM 01C27F01-01EN

4) Use a socket driver (width across flats, 5.5 mm)

to loosen the two bosses.

5) Carefully pull the CPU assembly straight

forward to remove it.

6) Disconnect the flat cable (cable with white

connector at the end) that connects the CPU

assembly and the capsule.

NOTE

Be careful not to apply excessive force to the

CPU assembly when removing it.

■ Mounting the CPU assembly
1) Connect the flat cable (with white connector)

between the CPU assembly and the capsule.

2) Connect the power cable (with brown

connector) to the CPU assembly.

NOTE

Make certain that the cables do not get pinched

between the case and the edge of the CPU

assembly.

3) Align and engage the zero-adjustment screw

pin with the groove on the bracket on the CPU

assembly. Then insert the CPU assembly

straight onto the post in the amplifier case.

4) Tighten the two bosses. Mount the RF

assembly, and the integral indicator (refer to

subsections 9.4.1 and 9.4.2).

IMPORTANT

Confirm that the zero-adjustment screw pin is

placed properly in the groove on the bracket prior

to tightening the two bosses. If it is not, the zero-

adjustment mechanism will be damaged.

5) Replace the cover.

9.4.4 Cleaning and Replacing the Capsule

Assembly

CAUTION

Precautions for the intrinsic safety explosion

prevention type instrument

Modification is not permitted by the user for

intrinsic safety explosion prevention type

transmitter. Consult our company when you want

to exchange capsule assembly.

Execute it only at time when the following points

were confirmed for exchange capsule assembly

in the same range of measurement.

• Exchanged capsule assembly uses the one

of the same specification.

• Tighten Setscrew to fix the transmitter

section and the pressure detector section

surely after ending maintenance.

This subsection describes the procedures for

cleaning and replacing the capsule assembly. (See

figure 9.3.)

■ Removing the Capsule Assembly

IMPORTANT

Exercise care as follows when cleaning the

capsule assembly.

• Handle the capsule assembly with care, and

be especially careful not to damage or distort

the diaphragms that contact the process

fluid.

• Do not use a chlorinated or acidic solution for

cleaning.

• Rinse with clean water after cleaning, please

dry until completely dry.

1) Remove the CPU assembly as shown in

subsection 9.4.3.

2) Remove the five setscrews, the stopper bolt,

and the stopper that connect the transmitter

section and pressure-detector section.

3) Remove the hexagon-head screw and the

stopper.

4) Separate the transmitter section and pressure-

detector section.

5) Clean the capsule assembly or replace with a

new one.