<9. Maintenance>

9-2

IM 01C27F01-01EN

Table 9.1

Instruments Required for Calibration

Name

Yokogawa-recommended Instrument

Remarks

Provisioning

device tool

• FieldMate (R2.02.01 or later)

• Provisioning Device Tool

• Infrared Adapter certified by Yokogawa

Supplier: ACTiSYS

Product name: IrDA InfraRed USB Adaptor

Product number: IR224UN

Field wireless

configuration

tool

• Field Wireless Integrated Gateway attached Software

Field Wireless Configurator

Field Wireless Management Tool

• Field Wireless System related Product

Plant Resource Manager (PRM) (R3.05 or later)

• Device Configuration Tool via Wireless Communication

FieldMate (R2.02.01 or later)

DeviceFile (R3.01.01 or later)

• Device Configuration Tool via Infrared Communication

FieldMate (R2.03.00 or later)

DeviceFile (R3.02.01 or later)

Digital

manometer

Model MT220 precision digital manometer

1) For 10 kPa class

Accuracy: ±(0.015% of rdg + 0.015% of F.S.). . . . for 0 to 10 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -10 to 0 kPa

2) For 130 kPa class

Accuracy: ±0.02% of rdg. . . . . . . . . . . . . . . . . . . . for 25 to 130 kPa

±5digits. . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 25 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

3) For 700 kPa class

Accuracy: ±(0.02% of rdg + 3digits) . . . . . . . . . . . for 100 to 700 kPa

±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 100 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

4) For 3000 kPa class

Accuracy: ±(0.02% of rdg + 10 digits). . . . . . . . . . for 0 to 3000 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

5) For 130 kPa abs class

Accuracy: ±(0.03% of rdg + 6 digits) . . . . . . . . . . . for 0 to 130 kPa abs

Select a manometer having

a pressure range close to

that of the transmitter.

Pressure

generator

Model MC100 pneumatic pressure standard for 200 kPa {2 kgf/cm

2

}, 25

kPa {2500 mmH

2

O}

Accuracy: ±0.05% of F.S.

Requires air pressure

supply.

Dead weight gauge tester 25 kPa {2500 mmH

2

O}

Accuracy: ±0.03% of setting

Select the one having a

pressure range close to that

of the transmitter.

Pressure

source

Model 6919 pressure regulator (pressure pump)

Pressure range: 0 to 133 kPa {1000 mmHg}

Prepare the vacuum pump

for negative pressure

ranges.

Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and

management procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1%

or higher level, there may be difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa

representatives from which the instrument was purchased or the nearest Yokogawa office.