Handling cautions, 1 model and specifications check, 2 unpacking – Yokogawa EJX530B User Manual

Page 9: 3 storage, Handling cautions -1, Model and specifications check -1, Unpacking -1, Storage -1

Advertising
background image

<2. Handling Cautions>

2-1

IM 01C27F01-01EN

2. Handling Cautions

This chapter provides important information on how

to handle the transmitter. Read this carefully before

using the transmitter.
EJX Series transmitters are thoroughly tested at the

factory before shipment. When taking delivery of an

instrument, visually check them to make sure that

no damage occurred during shipment.
Also check that all transmitter mounting hardware

shown in figure 2.1 is included. If the transmitter

is ordered without the mounting bracket and the

process connector, the transmitter mounting

hardware will not be included. After checking the

transmitter, carefully repack it in its box and keep it

there until you are ready to install it.

U-bolt Nut (L type)

Mounting bracket

U-bolt (L type)

U-bolt (S type)

U-bolt Nut

(S type)

F0201.ai

The antenna is a detachable type when

Amplifier housing code 8 is selected,

and no antenna is provided for Amplifier

housing code 9.

Antenna

Figure 2.1

Transmitter Mounting Hardware

2.1 Model and Specifications

Check

The model name and specifications are written on

the name plate attached to the case.

: Refer to USER'S MANUAL.

Made in Japan

TOKYO 180-8750 JAPAN

MODEL

SUFFIX

SUPPLY

OUTPUT

MWP

mA DC

V DC

STYLE

CAL

RNG

NO.

F0202.ai

Figure 2.2

Name Plate

2.2 Unpacking

Keep the transmitter in its original packaging to

prevent it from being damaged during shipment.

Do not unpack the transmitter until it reaches the

installation site.

2.3 Storage

The following precautions must be observed when

storing the instrument, especially for a long period.
(a) Select a storage area which meets the following

conditions:

• It is not exposed to rain or subject to water

seepage/leaks.

• Vibration and shock are kept to a minimum.

• It has an ambient temperature and relative

humidity within the following ranges.

Ambient temperature:

–40 to 85°C

–30 to 80°C LCD visible range

Relative humidity:

0% to 100% R.H.

Preferred temperature and humidity:

approx. 25°C and 65% R.H.

(b) When storing the transmitter, repack it carefully in

the packaging that it was originally shipped with.

(c) If the transmitter has been used, thoroughly

clean the chambers inside the cover flanges, so

that there is no process fluid remaining inside.

Before placing it in storage, also make sure that

the pressure-detector is securely connected to

the transmitter section.

(d) Preferably remove the batteries for storage. For

maximum battery life, the storage temperature

should not exceed 30°C.

NOTE

When storing the instrument with a battery

pack, it is recommended to put the instrument in

Deep Sleep mode to conserve the batteries. For

details on how to switch to Deep Sleep mode,

refer to subsection 8.3.12 “Switching to Deep

Sleep Mode”.

Advertising