4 disassembly and reassembly, Disassembly and reassembly -3, Caution – Yokogawa EJX530B User Manual

Page 62: Important

Advertising
background image

<9. Maintenance>

9-3

IM 01C27F01-01EN

P

Reference pressure

Pressure generator

Supply pressure

P

Reference pressure

Precision digital manometer

Pressure source

Using pressure

generator

Using pressure source

with manometer

F0901.ai

Figure 9.1

Instrument Connections

9.4 Disassembly and

Reassembly

CAUTION

Precautions for the intrinsic safety explosion

prevention type instrument
Intrinsic safe type transmitters must be, as

a rule, removed to a non-hazardous area

for maintenance and be disassembled and

reassembled to the original state. Check and

confirm the insulation when it is reassembled to

the original state.

Check and confirm the insulation when it is

reassembled to the original state.

Refer to section 2.7 “Insulation Resistance and

Dielectric Strength Test” for details of Resistance

Test.

IMPORTANT

• Perform the provisioning when replacing the

RF assembly. Refer to 7.4 Connecting to the

Field Wireless Network for details.

• Replace the batteries and perform the

parameter settings when replacing the CPU

assembly. Refer to 9.4.6 Replacing the

Batteries.

This section describes procedures for disassembly

and reassembly for maintenance and component

replacement.

CAUTION

Always remove Battery pack and shut off

pressures before disassembly and assembly.

Use proper tools for all operations.

Table 9.2 shows the tools required.

Table 9.2

Tools for Disassembly and Reassembly

Tool

Quantity

Remarks

Phillips

screwdriver

1

JIS B4633, No. 2

Slotted

screwdriver

1

Allen wrenches

3

JIS B4648

One each, nominal 3,

4 and 2.5 mm Allen

wrenches

Wrench

1

Width across flats, 17 mm

Torque wrench

1

Adjustable

wrench

1

Socket wrench

1

Width across flats, 16 mm

Socket driver

1

Width across flats, 5.5 mm

Tweezers

1