Data Cal 2000 Hardware Setup • 16

3. Place the docking station on the flat surface with enough room to

the right of the docking station (when viewed from the front) to fit all
of the docking modules side by side.

4. Place the first docking module to the right of the docking station

and line up the three tabs on the left side of the module with the
three recesses on the right side of the docking station.

5. Push the module toward the docking station so the tabs mate with

the recesses and the module is flush against the docking station. If
only one module is used in your system, then mechanical assembly
is done. If more modules are used in your system, continue with
step 6.

6. Place the next docking module to the right of the installed docking

module and line up the three tabs on the left side of the module with
the three recesses on the right side of the installed module.

7. Push the new module toward the installed module so the tabs mate

with the recesses and the new module is flush against the installed
module.

8. Repeat steps 6 & 7 for each additional docking module.

CAUTION: Make sure that the last module in the system, the

one on the far right, has the terminator switch set to
the down position. If the last module does not have
the terminator switch set to the down position, the
Data Cal 2000 will not operate properly.

Cable Installation

1. Install the RS-232C cable from the PC connector on the back of the

docking station to the serial port on a computer.

2. Plug the power cord into the power socket on the back of the

docking station. Do not connect the cord to an AC outlet at this
time.

3. Install a RS-485 cable from the MODULE connector on the back of

the docking station to the right (when viewed from the back) DB-15
connector on the back of the first module. If only one module is