35 • Setting Up the Data Cal 2000 Software

7. For each additional gas in the cylinder being entered move to the

next Symbol(Unit) field and repeat steps 4 - 6.

8. If the cylinder will be used for bump testing, click in the Bump test

selection box to place a check in it.

9. If the cylinder will be used for calibrating, click in the Calibration

selection box to place a check in it.

10. Click the OK button. A confirmation window appears.

11. Click Yes to save the new cylinder’s settings and return to the

Cylinders window.

12. Click the Close button to return to the main window.

Updating Existing Cylinder Parameters

1. From the Cylinders window, click on the cylinder in the cylinder list

that you want to update. It will be highlighted.

2. Click the Edit button. The CylinderDetail window for that cylinder

will appear.

Figure 20: Cylinder Detail Window, Editing a Cylinder