21 • Installing the Data Cal 2000 Docking Station Software

1. If the area around the calibration station is not considered a fresh

air area, an area free of combustible and toxic gases and of normal
oxygen content, 20.9%, install a tube on the AIR fitting on the back
of the docking station and route it to a fresh air area or connect a
cylinder of zero air to the fitting using a demand flow regulator.

2. Install the demand flow regulator(s) on the calibration gas

cylinder(s).

3. Connect the demand flow regulator for each calibration gas

cylinder that will be used with the docking station to the desired gas
port, GAS1, GAS2, GAS3, GAS4, or GAS5. These ports must be
defined in the docking station software during the software setup.
See “Setting Up the Solenoids” on page 36.

Installing the Data Cal 2000 Docking
Station Software

1. Launch Windows

®

.

2. Exit from all applications and open windows.

3. Insert the Data Cal 2000 Docking Station Software Installation CD

into your computer’s CD-ROM drive.

4. The Docking Station InstallShield Wizard comes up to guide you

through installation. Click Next to proceed to the License
Agreement window.

GX-2009
LEL/Oxy/H2S/CO

4-gas mix with CH4/Oxy/H2S/CO

GX-2009
LEL/Oxy/CO

3-gas mix with CH4/Oxy/CO

GX-2009
LEL/Oxy/H2S

4-gas mix with CH4/Oxy/H2S/CO

Table 2: Recommended Gas Cylinders

Typical Instrument Types

Recommended Calibration Gas

Cylinder(s)