Setting Up the Data Cal 2000 Software • 36

3. Update the cylinder parameters as desired.

4. Click the OK button. A confirmation window appears.

5. Click Yes to save the changes and return to the Cylinders window.

6. Click the Close button to return to the main window.

Setting Up the Solenoids

The solenoids are located inside the Data Cal 2000 docking station. They
control which fitting on the back of the docking station is used to draw
gas from a connected cylinder during a bump test or calibration. In the
case of the AIR fitting, it may draw from a zero air cylinder connected to it
or from the ambient air. Which fitting is used to draw gas at a particular
time during a calibration or bump test is determined by the software
depending on the solenoid cylinder assignment and the version of the
instrument that is being bump tested or calibrated by the Data Cal 2000.
To enter the Solenoids window, click the Solenoids button in the main
program window.

When the software is first installed, the solenoids are assigned as
follows:

the AIR solenoid controls the AIR fitting and it is assigned cylinder
number 0000 which is defined as “air”. This assignment is fixed and
cannot be changed.

Figure 21: Solenoids Window