19 • Data Cal 2000 Hardware Setup

CAUTION: The maximum recommended length for the

exhaust tube is 10 feet. Do not use more than 10
feet of tubing for the exhaust tube or bump test and
calibration accuracy will be adversely affected. The
tube that is shipped with the Docking Station is 10
feet long.

GX-2009

The DM-2009 only has one exhaust fitting instead of two like the
DM-2001 and DM-2003. Each DM-2009 in a Data Cal 2000 system
must have a dedicated exhaust tube.

Install the 5/16” ID flexible tube that is included with the DM-2009
on the exhaust fitting port and route the tube to an area where the
calibration gas can be safely exhausted.

CAUTION: The maximum recommended length for the

exhaust tube is 10 feet. Do not use more than 10
feet of tubing for the exhaust tube or bump test and
calibration accuracy will be adversely affected. The
tube that is shipped with the Docking Station is 10
feet long.

MODULE

PPC (RS-232C)

Route to open area

where calibration gas

can be exhausted.

Tubing, User Supplied

GAS OUT

AIR

GAS 1

GAS 2