RKI Instruments Data Cal 2000 User Manual

Page 81

Advertising
background image

Instrument/Module Function Menu • 76

document.

3. Click the OK button in the lower right corner of the Instrument

Information window.

4. A confirmation box will appear. Click the Yes button.

5. The software will save the memo in the instrument’s Memo folder in

the Logs Window and return to the main program window.

Download Function

This function can be activated by either right clicking an instrument that is
connected to the Data Cal 2000 and selecting Download(D) when the
pull down menu appears or by selecting an instrument that is connected
to the Data Cal 2000 and then using the alt/D key combination. The
Download function downloads all logged data and calibration information
from a GX-2003 and a GX-2009 and all logged data from a GX-2001
connected to the Data Cal 2000. If you select Download(D), a
confirmation box appears. Click the OK button to proceed with
downloading from the instrument.

Clear Logs Function

This function can be activated by either right clicking an instrument that is
connected to the Data Cal 2000 and selecting Clear Logs(C) when the
pull down menu appears or by selecting an instrument that is connected
to the Data Cal 2000 and then using the alt/C key combination. The Clear
Logs function clears all logged data and calibration files from an
instrument connected to the Data Cal 2000. If you select Clear Logs(C),
a confirmation box appears. Click the OK button to proceed with clearing
all the data from the instrument.

Reset Function

This function can be activated by either right clicking an instrument that is
connected to the Data Cal 2000 or a docking module and selecting
Reset(R) when the pull down menu appears. It can also be activated by
selecting an instrument that is connected to the Data Cal 2000 or a
docking module and then using the alt/R key combination. The Reset
function resets the selected docking module. This is useful if
communication errors occur with the docking module or the module does
not communicate with the Data Cal 2000. If you select Reset(R), a
confirmation box appears. Click the OK button to proceed with resetting
the selected docking module.