Figure 13: exhaust fitting connections – RKI Instruments Data Cal 2000 User Manual

Page 23

Advertising
background image

Data Cal 2000 Hardware Setup • 18

3. Repeat step 2 for each successive module in the system until you

get to the last module.

4. Install the black fitting plug on the left CAL. GAS fitting of the last

module in the system.

Module Exhaust Fitting Connections

GX-2001 and/or GX-2003

1. Install a clear fitting plug, supplied with the system, on the bottom

port of the exhaust fitting of the first module in the system.

2. Install a clear tube, supplied with the system, between the top

exhaust fitting port on the first module and the bottom exhaust
fitting port on the next module.

3. Repeat step 2 for each successive module in the system until you

get to the last module.

4. On the last module in the system, install the 5/16” ID flexible tube

that is included with the Docking Station on the top exhaust fitting
port and route the tube to an area where the calibration gas can be
safely exhausted.

MODULE

PC(RS-232C)

GAS OUT

AIR

GAS1

GAS2

Install plug over fitting.

Tubing,Supplied with System

Tubing,
User Supplied

Route to open area

where calibration gas

can be exhausted.

Figure 13: Exhaust Fitting Connections