49 • Charging Instruments With The Data Cal 2000

2. The LED on the left front of the docking module will turn off. The

LED on the right front of the module will start blinking yellow.

3. If the battery pack is fully charged, then the blinking yellow LED will

turn solid green in a few minutes.

If the battery pack is drained enough for the module to charge it,
the right LED will continue to blink yellow while charging is taking
place. The GX-2001 and GX-2003 each take approximately 90
minutes to charge while the GX-2009 will take approximately 3
hours to fully charge.

NOTE: If the GX-2009 is left in the DM-2009 for a long period of

time, the dock pump will run for 12 seconds every 2
minutes in order to keep excess hydrogen generated
from the NiMH batteries from affecting the CO sensor.

To recharge the batteries in an instrument without launching the Data Cal
2000 Docking Station Software:

1. Turn on the docking station. The right LED on the docking module

will start blinking green.

2. Install the instrument into the docking module. If only charging is

being performed, it is not necessary to install the sample adapter to
a GX-2001 or the sample tubes to a GX-2003. Since the GX-2009
does not require any attachments, it can be installed in the
DM-2009 normally. The right LED on the docking module will start
blinking yellow.

3. If the battery pack is fully charged, then the blinking yellow LED will

turn solid green in a few minutes.

If the battery pack is drained enough for the module to charge it,
the right LED will continue to blink yellow while charging is taking
place. The GX-2001 and GX-2003 each take approximately 90
minutes to charge while the GX-2009 will take approximately 3
hours to fully charge.