Logs Window • 58

Calibration Data

The GX-2003 saves the calibration information for the last 20 calibrations
while the GX-2009 saves the calibration information for the last 100
calibrations. When an autocalibration is performed, the calibration results
are saved as a file. When single calibration is used, the GX-2003 and
GX-2009 consider each time you calibrate a channel a calibration and a
file is saved for that calibration. Once a GX-2003 or a GX-2009 have
reached their maximum number of calibration files, the oldest calibration
file is overwritten when a new one is saved.

When a GX-2003 or a GX-2009 is connected to the Data Cal 2000, the
calibration files in it are retrieved and stored by the software. In addition,
all calibrations performed by the Data Cal 2000 on a GX-2003 or a
GX-2009 are stored by the software.

A GX-2001 only saves calibration information for the most recent
successful calibration for each channel. When a GX-2001 is connected
to the Data Cal 2000, this information is not retrieved by the Data Cal
2000. However, all calibrations performed by the Data Cal 2000 on a
GX-2001 are stored by the software.

To view and perform desired operations with the calibration files:

1. While in the Logs window, open folders in the left side of the

window until the Calibration folder you wish to view is visible.

2. If you are viewing data in base view format with the data organized

by type, expand the Calibration folder. Folders organized by year/
month will appear below the Calibration folder. Click on the folder
whose contents you wish to see. The calibration files in the folder
will be shown in the upper right side of the Logs window.