Diffusion Eagle Instruction Manual

Operation • 13

NOTE: If the Remote detector has been disconnected from the Eagle

for more than a few minutes, such as during shipment, wait
15 minutes after connecting the Remote detector before
turning on the Eagle to allow the CO and H

2

S sensors to

stabilize.

To attach the Remote Detector using an extender cable, carefully fit
the end of the extender cable that has a coupling to the Eagle’s
connector. Make sure to line up the tabs inside the coupling with the
matching notches in the cable connector as shown below. Then line
up the tab on the extender cable connector with the notch inside the
connector on the Eagle. Gently push the extender cable toward the
Eagle. Turn the coupling clockwise on the extender cable to secure
the cable to the Eagle’s bottom case assembly. Next, carefully fit the
other end of the extender cable to the connector on the Remote
Detector. Position the Remote Detector coupling in the unlocked
position as shown above. Then line up the tab inside the Remote
Detector connector with the notch inside the connector on the Eagle.
Gently push the extender cable toward the Remote Detector. Tighten
the coupling on the Remote Detector connector to secure the extender
cable to the Remote Detector.

Rotate to Lock

Align Keys in Key

Way For Assembly

Cable End
Connector

Eagle

Connector