Diffusion Eagle Instruction Manual

Setup Mode • 45

To reset all default alarm point settings:

1. From the main menu, select the DEFAULT menu option. The Set Default

All screen displays.

2. Press the DISP/ADJ button to display the Set Default Alarm screen.

3. Press the AIR/

button to reset all alarm points to their default settings.

The messages SAVING DATA and END display, then the main menu
displays.

To reset the oxygen zero setting:

1. From the main menu, select the DEFAULT menu option. The Set Default

All screen displays.

2. Press the DISP/ADJ button. The Set Default Alarm screen displays.

3. Press the DISP/ADJ button to display the Set Default Oxygen Zero

screen.

4. Press the AIR/

button to reset the oxygen zero setting to its default

value. The messages SAVING DATA and END display, then the main
menu displays.