Updating the lunch break setting, Updating the alarm latching setting – RKI Instruments Eagle Series User Manual

Page 43

Advertising
background image

38 • Setup Mode

Diffusion Eagle Instruction Manual

Updating the Lunch Break Setting

With Lunch Break OFF (default), the Eagle automatically starts new TWA
and PEAK reading collection at start up.

With Lunch Break ON, the RESUME screen displays during start up. From
this screen, you can choose to continue accumulating TWA and PEAK
readings from the last time the Eagle was used or start collecting new
readings.

1. From the main menu, select the LUNCH BREAK menu option.

2. Press the AIR/

or SHIFT/

button to display the desired setting.

3. Press the POWER/ENTER button to enter the setting and return to the

main menu.

Updating the Alarm Latching Setting

With Alarm Latching ON, the Eagle remains in alarm condition until the
alarm condition passes and the RESET/SILENCE is pressed.

With Alarm Latching OFF, the Eagle automatically resets its alarm when the
alarm condition passes.

1. From the main menu, select the ALARM LATCHING menu option.

2. Press the AIR/

or SHIFT/

button to display the desired setting.

3. Press the POWER/ENTER button to enter the setting and return to the

main menu.