RKI Instruments Eagle Series User Manual

Page 35

Advertising
background image

30 • Setup Mode

Diffusion Eagle Instruction Manual

5. Press the POWER/ENTER button to enter the new target gas label.

A screen displays that shows the full-scale PPM setting, which
corresponds to 100% LEL, and display increments for the target gas label
you selected. If you select *** as the gas label, you must update the full-
scale PPM setting to correspond to 100% LEL for the target gas.

The number in parenthesis indicates the display increment for that
portion of the PPM range. In the example below, the PPM reading would
display in increments of:

5 from 0 to 100 ppm

50 from 100 to 1000 PPM and 1000 to 10,000 PPM

250 from 10,000 to 50,000 PPM

If you entered a label other than ***, continue with step 6. If you entered ***,
go to the next section, “Setting a custom target gas label.”

6. Press the POWER/ENTER button to return to the Gas Combinations

menu.

7. To exit the Gas Combinations menu, press the SHIFT/

button until the

prompt is next to Channel 4, then press the SHIFT/

button again. The

ESCAPE message displays.

8. Press the POWER/ENTER button. The message SAVING DATA

displays, then the main menu displays.

Setting a custom target gas label

1. With the prompt next to the target gas label setting (***), press the

POWER/ENTER button. The first asterisk flashes.

2. Press the AIR/

and SHIFT/

buttons to display the desired character.

Available characters are A through Z, 0 through 9, and a blank space.

* * *

5

0

(

P

0

0

>

0 0

P M

1

0

5

)

0

1

0 0 0

(

5

)

0

0

1

0 0

(

5

)

0

5

0

0

0 0

(

5

)

0

2

Advertising