Diffusion Eagle Instruction Manual

Alarms • 21

Low battery alarm

When the battery charge drops near the lower limit, the Eagle displays the
following screen (BAT flashes). For alkaline batteries, you have
approximately 3 hours of use remaining; for Ni-Cd batteries you have
approximately 15 minutes of use remaining.

When the battery voltage drops to the minimum limit, the following screen
displays, the alarm lights are on continuously, and the buzzer sounds a steady
tone. The Eagle is not operational as a gas monitoring device when this
screen displays.

NOTE: If you are using the optional data logging accessory and the Eagle

goes into Low Battery ALARM, shut off the Eagle in order to save
the current data logging session.

Sensor failure alarm and emergency operation

The Eagle continuously monitors itself for proper operation. If a malfunction
occurs, the Eagle alerts you with audible and visual alarms.

If a sensor fails during start-up or normal operation:

The message FAIL SENSOR displays.

The failed sensor displays in parenthesis.

The buzzer sounds a steady tone.

The alarms lights flash.

C H

4

O

Y

C O

H

2

S

X

2 0 . 9

0

0

0

.

0

L

V
P
P

P
P

M
M

E

O

L
L

%
%

B
A

T

Low Battery

Warning

B A T T E R Y

M I N .

4 . 5 V

B A T T E R Y

G

C H

N

A

E

Low Battery

ALARM

> < O

Y

>

<

X

> <

<

>

F A I L

S E N S O R