Diffusion Eagle Instruction Manual

Setup Mode • 33

Updating the full-scale setting

1. Press the SHIFT/

button to place the prompt in the second line, then

press the POWER/ENTER button to update the full-scale setting. The
full-scale setting flashes.

The maximum full-scale setting for the oxygen channel is 40.0 VOL%;
the minimum setting is 25.0 VOL%. The default setting is 40.0 VOL%.

2. Press the AIR/

and SHIFT/

buttons to display the desired full-scale

setting, then press the POWER/ENTER button to enter the setting. The
prompt flashes.

Updating the display increment setting

1. Press the SHIFT/

button to place the prompt in the third line, then press

the POWER/ENTER button. The display increment setting flashes. The
allowable settings are 0.2 VOL% (default) and 0.5 VOL%.

2. Press the AIR/

or SHIFT/

button to display the desired display

increment setting, then press the POWER/ENTER button to enter the
setting. The prompt flashes.

Returning to the main menu

1. Press the SHIFT/

button. The ESCAPE message displays. (Press the

AIR/

button to return to the previous screen.)

2. Press the POWER/ENTER button to save the new settings. The OTHER

GAS SET message displays, then the Gas Combinations menu displays.

3. To exit the Gas Combinations menu, press the SHIFT/

button until the

prompt is next to Channel 4, then press the SHIFT/

button again. The

ESCAPE message displays.

4. Press the POWER/ENTER button. The message SAVING DATA

displays, then the main menu displays.

Updating toxics channel settings

This section describes how to update the target gas label, set a custom gas
label, and update the full-scale and display increment settings for a toxic gas
channel.