18 • Operation

Diffusion Eagle Instruction Manual

Setting User Access

The CAL/SETUP switch controls the Eagle functions available to the user.
The switch setting does not affect the Eagle’s ability to display gas readings
and indicate gas and malfunction alarms.

1. Turn off the Eagle.

2. Unscrew the two large screws on the top of the case.

3. Turn over the top half of the case, and locate the CAL/SETUP switch

(SW2) near the middle along one edge of the main processor board.

CAUTION:

The Methane Elimination switch (SW1) is on the opposite
edge of the board near the front end. DO NOT confuse
these two switches.

4. Place the CAL/SETUP switch in the appropriate position.

To give the Eagle access to all modes, place the switch in the ON
position.

To limit the Eagle to normal operating and display modes, place the
switch in the OFF position. (The Eagle prevents access to the setup
and calibration modes by disabling the SHIFT/

button.)

5. Place the top of the case in its original position, then secure it with the

large screws you loosened in step 2.

6. Turn on the Eagle.

NOTE: Make sure the Eagle’s calibration is current and the setup options

appropriate and safe for the operating environment before placing
the CAL/SETUP switch in the OFF position.

Turning Off the Eagle

To turn off the Eagle, press and hold down the POWER/ENTER button until
GOOD-BYE displays, then release the button. (You must wait for GOOD-
BYE to disappear before you can turn on the Eagle again.)