Updating the auto calibration settings – RKI Instruments Eagle Series User Manual

Page 45

Advertising
background image

40 • Setup Mode

Diffusion Eagle Instruction Manual

Updating the Auto Calibration Settings

The Eagle stores default calibration settings. This allows you to calibrate all
Eagle channels simultaneously with a calibration cylinder that contains all
required target gases (for example the RKI four-gas calibration cylinder).

The Eagle includes default auto calibration settings for most target gases. For
gases without default auto calibration, the setting is 0.

NOTE: You can also update auto calibration settings in Calibration mode. If

you update auto calibration settings in Calibration mode, you must
continue with the calibration procedure. Updating these settings in
Setup mode allows you to update the settings without calibrating the
sensors.

1. From the main menu, select the AUTO CALIBRATION menu option. (To

display the combustible gas channel in PPM, press the LEL/PPM button.)

2. Press and hold the SHIFT/

button, then press the DISP/ADJ button. The

Auto Calibration screen for the combustible gas channel displays.

3. Press the AIR/

or SHIFT/

button to display the desired setting.

4. Press the POWER/ENTER button to enter the new setting. The Auto

Calibration screen for the next channel displays.

5. Repeat steps 4 and 5 for the remaining channels. (You must scroll

through all remaining channels before you can exit the Auto Calibration
screen.)

6. When the Auto Calibration screen for the last channel displays, press the

POWER/ENTER button to return to the main Auto Calibration screen.

7. Press the POWER/ENTER button to return to the main menu.