Diffusion Eagle Instruction Manual

Maintenance • 53

Replacing the combustibles sensor

Replace the combustibles sensor when:

The combustibles channel cannot be calibrated correctly.

The LEL reading cannot be set to 0 by the Demand Zero command.

To replace the combustibles sensor:

1. Take the Eagle to a non-hazardous location, and turn the power off.

2. Remove the Remote Detector from the Eagle.

3. On the Remote Detector, unscrew the four screws on top of the case using

a flat blade screwdriver. Carefully lift the top of the case and place it to
one side. Be careful not to pull hard on the top because a two-wire cable
connects the top to the circuit board inside the case.

H2S

Sensor

Charcoal

Filter

To CN4

Combustibles

Sensor

To CN3

CO Sensor

To CN2

To CN1

Filter Boot

Oxygen Sensor