Updating the zero follow settings – RKI Instruments Eagle Series User Manual

Page 48

Advertising
background image

Diffusion Eagle Instruction Manual

Setup Mode • 43

Updating the Time Calibration Setting (data log option)

This setting indicates how often the Eagle alerts you to needed calibration.
The minimum setting is 1 day; the maximum setting is 180 days. The default
setting is “off”.

Tip: Press and hold the AIR/

or SHIFT/

button to rapidly scroll through

settings.

1. From the main menu, select the TIME CALIBRATION menu option.

2. Press the AIR/

and SHIFT/

buttons to display the desired setting.

3. Press the POWER/ENTER button to enter the setting and return to the

main menu.

Updating the Date and Time Settings (data log option)

The Data Logger uses the data and time to identify entries.

1. From the main menu, select the DATE/TIME menu option.

2. Press the AIR/

or SHIFT/

button to display the desired month.

3. Press the POWER/ENTER button to enter the setting. The day setting

flashes.

4. Repeat steps 2 and 3 to enter the day, year, hours, and minutes settings.

The main menu displays after you enter the minutes setting.

Updating the Zero Follow Settings

The Zero Follow setting is not intended for customer setup. The default
setting for most target gases is ON. The default setting for carbon dioxide
sensors and some configurations of non-standard toxic gas sensors is OFF.
The oxygen sensor does not include this feature.

CAUTION:

Contact RKI, Instruments Inc., before changing this
setting.

L

T

A

A

C

I

Y

T

E

S

D

T I M E

A

R

I O

B

R E Q U

S T

E

N

S

f

f

o

P

A

1 7

1

4 :

R

9 9 8

1

3 0

Advertising