Maintenance • 51

Maintenance

Displaying the Battery Voltage

Check the battery voltage periodically. Replace or recharge the batteries
before the voltage drops to 4.5 V.

WARNING: Take the Eagle to a non-hazardous location before replacing or

recharging the batteries.

To display the battery voltage:

1. From the normal screen, press the DISP/ADJ button to enter display

mode.

2. Press the DISP/ADJ button until the Battery Voltage screen displays.

3. To exit display mode, press the DISP/ADJ button until the normal screen

displays.

Replacing Alkaline Batteries

1. Turn off the Eagle.

2. Unscrew the two large screws on the top of the case, then carefully lift the

top of the case and lay it aside.

3. Remove the batteries and verify that the battery compartment and

electrical contacts are clean.

4. Insert fresh batteries following the label on the inside of the case.

5. Place the top of the case in its original position, then secure it with the

large screws you loosened in step 1.

Recharging Ni-Cd Batteries

1. Turn off the Eagle.

2. Plug the cord from the charger into the Eagle’s battery charger connector.

3. Plug in the AC line cord or 12 VDC supply to the charger. A full charge

takes approximately 8 to 12 hours.

4. Unplug the supply and the charger before using the Eagle. See the

charger label for directions.

WARNING: Use only the Eagle battery charger to charge batteries. The

B A T T E R Y