Diffusion Eagle Instruction Manual

Setup Mode • 31

3. Press the POWER/ENTER button to enter the displayed character. The

next character flashes.

4. Repeat steps 2 and 3 to enter the remaining characters. When you enter

the last character, the prompt flashes.

Updating the full-scale PPM setting

CAUTION:

The full-scale PPM setting must correspond to 100% LEL
for the target gas in order for the Eagle to display
accurate PPM readings for the combustible gas channel.

1. Press the SHIFT/

button to place the prompt in the second line, then

press the POWER/ENTER button to update the full-scale setting. The
full-scale setting flashes.

The maximum full-scale setting for the combustible gas channel is
50,000 PPM; the minimum setting is 1000 ppm. The default setting is
50,000 ppm.

2. Press the AIR/

and SHIFT/

buttons to display the desired full-scale

setting (see Table 5), then press the POWER/ENTER button to enter the
setting. The prompt flashes.

Table 5: full-scale PPM Readings

Equivalent to 100% LEL

Target Gas

full-scale Setting

Methane (CH

4

)

50,000 ppm

Hexane

11,000 ppm

Hydrogen

40,000 ppm

Pentane

15,000 ppm

Styrene

9,000 ppm

IPA

20,000 ppm

Isobutane

18,000 ppm

Propane

21,000 ppm

Propylene

20,000 ppm

Toluene

11,000 ppm

Ethane

30,000 ppm

Ethanol

33,000 ppm

Benzene

12,000 ppm