34 • Setup Mode

Diffusion Eagle Instruction Manual

Updating the target gas label

1. From the main menu, select the GAS COMBINATIONS menu option.

2. Press the POWER/ENTER button to display the Gas Combinations

menu.

3. Use the AIR/

or SHIFT/

button to place the prompt next to the toxic

gas channel (in this example H2S or CO).

4. Press the POWER/ENTER button. The toxic target gas label flashes. This

indicates that this setting can now be updated.

5. Press the AIR/

or SHIFT/

buttons to display the available target gas

labels for the toxic gas channel (H2S, CO, SO2, Cl2, NH3, CO2 (5.00%),
CO2 (10000 PPM), CO2 (5000 PPM), ***, and NOT USED).

6. Press the POWER/ENTER button to enter the new target gas label.

If you entered a label other than ***, continue with step 7. If you entered ***,
go to the next section, “Setting a custom target gas label.”

7. To exit the Gas Combinations menu, press the SHIFT/

button until the

prompt is next to Channel 4, then press the SHIFT/

button again. The

ESCAPE message displays.

8. Press the POWER/ENTER button. The message SAVING DATA

displays, then the main menu displays.

Setting a custom target gas label

1. With the prompt next to the target gas label setting (***), press the

POWER/ENTER button. The first asterisk flashes.

2. Press the AIR/

and SHIFT/

buttons to display the desired character.

Available characters are A through Z, 0 through 9, and a blank space.

3. Press the POWER/ENTER button to enter the displayed character. The

next character flashes.

4. Repeat steps 2 and 3 to enter the remaining characters. When you enter

the last character, the prompt flashes.