21 • Setting the Operational Parameters in Edit Mode

Setting the Bump Test Parameters

Do the following to set the bump test parameters.

1.

Turn on the SDM-2012 with an instrument and establish a connection between
them as described above in "Turning on the SDM-2012 with an Instrument".
The instrument will display the home screen.

2.

Press and hold the EDIT ENTER button for 2 seconds. The instrument will
display the Edit Mode screen.

3.

Press and release the BUMP

T

button. The instrument will alternate between

the following screens that display the four bump test parameters and their
settings. The BUMP

T

LED will turn green.

4.

If you wish to cancel setting the bump test parameters or were just viewing the
parameters to confirm their values, press and release the BUMP

T

button to

return to the home screen.
To continue and change parameters, press and release the EDIT ENTER
button. The Air Sample Time parameter will be displayed. For a description of
the Air Sample Time bump test parameter, see “Bump Test & Calibration