25 • Setting the Operational Parameters in Edit Mode
5.
Use the BUMP
T
and CAL
S
buttons to scroll through the different calibration
parameters.
6.
To edit a parameter, press and release the EDIT ENTER button when that
parameter is displayed. An “E” will appear to the right of the parameter name.
7.
Use the BUMP
T
and CAL
S
buttons to set the parameter to the desired
value, then press and release the EDIT ENTER button. The “E” to the right of
parameter name will disappear.
8.
Repeat step 6 - step 7 to set any other parameters.
9.
When you are done setting the parameters, use the BUMP
T
button to scroll
past the GAS--IN parameter screen. The screen will indicate that the
parameter changes have been saved.
The instrument will return to the home screen.
90
GAS--IN
45
E AIR--IN
45
90
DECISION