89 • Troubleshooting

• No IR connection

between
instrument and
SDM-2012

• SDM-2012 is not turned

on.

• The instrument is not

correctly inserted into the
instrument cradle.

• The IR window is dirty.

1. Turn on the SDM-2012. If it does

not turn on, check that the AC
adapter is plugged into an AC
socket and to the jack on the
back of the SDM-2012.

2. Check to make sure the

instrument is inserted properly.

3. Clean the IR window on the

SDM-2012.

4. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• Control buttons do

not function
correctly for
standalone
operation

• The SDM-2012 is

connected to a computer
and the PC Controller
program is running.

1. Disconnect the USB cable from

the back of the SDM-2012 if it is
connected. See the SDM-2012
PC Controlled Configuration
Operator’s Manual for
instructions to use the SDM-2012
in PC Controlled configuration.

• The COPY

function does not
copy files to the
USB drive

• The USB drive is

defective.

• The USB drive is not

inserted properly.

• The SDM-2012 is

connected to a computer,
the PC Controller program
is running, and an
instrument is turned on.

1. Perform a COPY operation with a

different USB drive.

2. Be sure that the USB drive is

inserted correctly.

3. Disconnect the SDM-2012 from

the computer.

4. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• The Li-ion battery

pack in the
instrument is not
charging

• The instrument has an

alkaline battery pack
installed.

• The battery pack is faulty.

• The instrument is not in

the proper recess of the
instrument cradle.

• The instrument is not

properly inserted in the
instrument cradle.

• The charger is

malfunctioning.

1. Check to make sure your

instrument doesn’t have alkaline
battery pack installed.

2. Replace the battery pack and

attempt to charge it.

3. Be sure that the instrument is

inserted in the front recess of the
instrument cradle.

4. Be sure that the instrument is

fully inserted in the instrument
cradle.

5. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.