Cal g2, Cal air, P p f p – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 59

Advertising
background image

Calibrating a GX-2012 • 54

menu item in Edit Mode is set to Std, continue with step 17.

If a %volume sensor is installed in your GX-2012 and if the GAS menu item in
Edit Mode is set to ALL, GAS 2 calibration gas will then be applied to that
sensor for the time defined by the GAS TIME calibration parameter.

17. The SDM-2012 will then purge the system with fresh air for the time defined by

the AIR TIME calibration parameter.

18. After the fresh air purge is complete, the screen will alternate between the gas

readings and the calibration results. If the calibration passed, the auto
calibration values will be displayed. If the calibration failed, the gas readings at
the end of the gas application will be displayed. The calibration results are
indicated with a P (pass) or an F (fail) to the left of the channel units. Use the
EDIT ENTER button to scroll between the calibration results for the standard
four sensors and the calibration results for the %volume sensor (if installed).
The CAL

S

LED will stop blinking and be steadily green if all channels passed

the calibration or steadily red if any channel failed the calibration. In the
examples below, the CO sensor failed calibration.

Figure 29: Calibration Results, Standard 4-Gas GX-2012 or

GAS Menu Item in Edit Mode Set to Std

98

vol%

CH

4

CAL G2


CAL AIR

P
P
F
P

%LEL

%

ppm

ppm

CH

4

O

2

CO

H

2

S

CAL

49
12.0
12
24.5

%LEL

%

ppm

ppm

CH

4

O

2

CO

H

2

S

CAL